Function Insulation monitor for medical areas
System type AC
Nominal voltage to 500 V
Reponse value type Einstellbar
Reponse value kOhm ... kOhm 50 - 500 Ohm
With auxiliary voltage +
Earth fault indicator +
Enclosure design Installationsverteiler
Width 52,5 mm
Type IN 5880/711
According to IEC/EN 61 557-8 For rooms used for medical purposesaccording to IEC 60364-7-710, DIN VDE 0100-710 For three-phase and A.C. power systems with 0 &hellip, 500 Vand 10 &hellip, 1000 Hz (IT power systems) Adjustable alarm value for ground fault RAL of 50 &hellip, 500 k&Omega, Measuring circuit with broken wire protection As option, programmable for storing or non-storing of errors With reset and test button Additional external reset and test buttons can be connected LED indicators for operation, insulation fault, and interruption ofMeasuring circuit 2 changeover contacts With LED chain for indication of the current insulation status IP 5880/711 for connection of the test and display panel UP 5862 52.5 mm width''